NEW! MT160
The MT160 thickness gauge is an inexpensive, easy to use thickness gauge
for quick spot checks of metals and most plastics. Selectable resolution of
0.01" or 0.001". Includes data storage of up to 20 files with 99
measurements per file.
Specifications
* Measuring Range in steel: 0.03 to 12.0 inches (0.75 to 300mm)
* Sound Velocity Range: 1000 to 9999 m/s
* Resolution: 0.01 inches (0.1mm) or 0.001 inches (0.01mm)
* Accuracy: ±0.5% of Thickness (0.04mm)
Depending on material and conditions
* Display: 4.5 digits LCD with EL backlight
* Units: Metric or Imperial (selectable)
* Four measurement readings per second for single point measurement,
and ten per second for Scan Mode
* Memory for up to 20 files (up to 99 values for each file) of stored values
* Power Source: Two “AA” size x 1.5 Volt alkaline batteries
100 hours typical operating time (EL backlight off).
* Physical Dimensions: 6H x 3W x 1.25D inches (150×74×32 mm)
* Weight: 9 oz. (245g)
* Supplied in padded nyon carry
bag with outside pocket and shoulder strap
* 1 year warranty
Click
for MT160 .pdf document
Transducer Selection
Model Freq Dia.
Measuring
Range Lower Limit
N02 2 MHz 22mm
3.0 to 300.0mm
20mm
For measurement of thick, highly attenuating, or
highly scattering materials
N05 5 MHz 10mm
1.2 to 230.0mm
20mm diameter x 3.0mm
For normal measurement applications - with top-exiting cable
N05/90 5 MHz 10mm
1.2 to 230.0mm ; 20mm diameter x
3.0mm
For normal measurement applications - with
right angle side-exiting cable
N07 7 MHz 6mm
0.75 to 80.0mm
15mm diameter x 2.0mm
For measurement of thin wall or small curvature pipe wall
HT5 5 MHz 14mm
3.0 to 200.0mm
30mm
For high temperature measurement (less than 570F / 300C)
Back to Top
NDT 715
The NDT
715 offers the same specifications as our field proven NDT-710 gauge. It is a small,
hand held thickness gauge, only 2.5"W x 4.75"H x 1.25"D and weighing only
10 oz. It is designed to make reliable spot check measurements an has a SCAN MODE
that will capture and display the thinnest measurement taken during a strip scan along a
length of material. Housed in a extruded, anodized aluminum housing with nickel
plated aluminum end caps and a weather resistant keypad the gauge is suitable for your
toughest working condition. The NDT 715 is also more resistant to potential
interference from RFI fields than plastic housed instruments.
Click
for NDT-715 .pdf document
Sound Velocity Range: 0.0492 to
0.3930 in/ms (1250 to 10,000 m/s)
Size: 1.7" H. x 2.4" W. x 4.75" L (32mm x 62mm x 120mm)
Temperature Range: - 20 to 120 F (-20 to 50 C)
Weight: 10 oz. with batteries (284 grams)
Powered by two AA 1.5V alkaline batteries
2 year warranty
SPECIFICATIONS
Principle:
Ultrasonic Pulse-echo with dual element transducer
Range: 0.030 (0.75mm) to 19.99 (199.99mm) depending upon transducer used and material
Resolution: 0.001 (0.01mm)
Automatic linearity adjustment
Automatic Zero adjustment with transducer on calibration block
Automatic Shut-Off 2 minutes after last reading or last key press
Built in Probe Zero calibration block
Easy to read back lighted Liquid Crystal Display (LCD) with Coupling Indicator
Instant switching between English or Metric measurement
Up to 75 hours continuous operation using alkaline batteries
Low Battery Indicator

Standard Kit Includes the NDT-715 Gauge with alkaline batteries, 1/4" element
diameter x 5.0 MHz Transducer on 4 foot long integral cable, 4 oz. bottle of ultrasonic
gel Couplant, Operation Manual and padded nylon Carry Bag with shoulder strap.
Optional foam lined plastic case.
Back
to Top
NDT 815 MVX
A very versatile digital thickness gauge with
thickness display, A-Scan and B-Scan display and Data Logging Capability.
Click for NDT-815 .pdf document

The NDT 815 is equipped with multiple viewing options to provide
users with a complete set of inspection tools: (RF waveform, +/- Rectified waveform, Time
based B-Scan, and Large Digits).
The
adjustable square wave pulser provides the flexibility necessary for both high resolution
and penetration requirements.
The A-Scan rectified mode is commonly used for
detecting flaws/pits in pulse-echo mode and measuring thru-paint and coatings in echo-echo
mode.
The time-based B-Scan feature of the NDT 815 displays a cross section of the
test material. It is commonly used to display the profile of the bottom surface of the
test material.
Built-in AGC gain control for thru-paint measurements
in multi-mode operation.
Prove up flaw detection capability when
using the optional single to dual transducer adapter cable along with an angle beam
transdcuer and wedge assembly.
The variety of calibration options is just one more
example of NDT 815’s overall versatility.
The NDT 815 has the ability to store
up to 64 custom user-defined setups. All factory setups can be selected, edited and saved
to any setup location.
The NDT 815 is equipped with an alpha-numeric data logger to
provide increased versatility for those custom reporting needs.
The built-in transducer types offer increased
linearity between transducers.
The high speed scan feature speeds up the
inspection process by making 32 measurements per second. Remove transducer from the test
material and display the minimum measurement scanned.
Use the visual alarm to set hi and lo limits
for applications requiring specific tolerances. If the actual thickness value is above or
below the limits, a red light is illuminated.
Use the find feature to locate the detection
point, while automatically adjusting the display to bring the signal into view.
Complete with Windows® based software and
interface cable for transferring data to & from a PC.
2 year limited warranty
SPECIFICATIONS
Physical
Weight: 13.5 ounces (with batteries).
Size: 2.5 W x 6.5 H x 1.24 D inches (63.5 W x 165 H x 31.5 D mm).
Operating Temperature: -14° to 140°F (-10° to 60°C).
Keyboard: Membrane switch with twelve tactile keys.
Case: Extruded aluminum body with nickel-plated aluminum end caps (gasket sealed).
Data Output: Bi-directional RS232 serial port. Windows® PC interface software.
Display: 1/8 in. VGA grayscale display (240 x 160 pixels). Viewable area 2.4 x 1.8
in. (62 x 45.7mm).
EL backlit (on/off/auto).
Ultrasonic
Measurement Modes: Pulse-Echo (flaws, pits) and Echo-Echo (thru-paint)
Pulser: Square wave pulser with adjustable pulse width (spike, thin, wide).
Receiver: Manual or AGC gain control with 40dB range, depending on mode selected.
Timing: 20 MHz with ultra low power 8 bit digitizer.
Power Source
Three 1.5V alkaline or 1.2V NiCad AA cells. Typically operates for 150 hours on alkaline
and 100 hours on NiCad (charger not included.)
Auto power off if idle 5 min. Battery status icon.
Measuring
Range in Pulse-Echo Mode: (Pit&Flaw Detection) measures from 0.025 to 9.999
inches
(0.63 to 254 millimeters).
Range in Echo-Echo Mode: (Thru Paint & Coatings) measures from 0.10 to 4.00
inches
(2.54 to 102 millimeters).
Range will vary +/- depending on the thickness of coating.
Resolution: +/- .001 inches (0.01 mm)
Velocity Range: .0492 to .3936 in./ms 1250 to 9999 meters/sec
Calibration: Single and Two point calibration option, or selection of basic material
types.
Units: English & Metric
Display
Display Views:
A-Scan Rectified +/- (flaw view) RF (full waveform view)
B-Scan Cross sectional view. Display speed of 15 secs per screen.
Large Digits Standard thickness view. Digit Height: 0.400 in (10mm).
Scan Bar 6 readings per second. Viewable in B-Scan and Large Digit views.
Repeatability Bar Graph Bar graph indicates stability of reading
Data Logger: 12,000 pages of memory (alpha numeric storage).
Page contents: 1 reading and 1 waveform per page.
OBSTRUCT to indicate inaccessible locations.
Memory: 16 megabit non-volatile ram.
Transducer
Transducer Types: Dual Element (1 to 10 MHz).
Locking quick disconnect “ 00” LEMO connectors. Standard 4 foot cable.
Custom transducers and cable lengths available for special applications.
Features
Setups: 64 custom user-definable setups. Factory setups can also be edited by
the user.
Gates: Single gate in pulse-echo mode, or single gate with holdoff in echo-echo
mode. Adjustable threshold. Selectable Transducers: Selectable transducer types with
built-in dual path error correction for improved linearity.
Alarm Mode: Set HI and LO tolerances with audible beeper and visual LEDs.
Fast-Scan Mode: Takes 32 readings per second and displays the minimum reading found
when the transducer is removed.
Certification
Factory calibration: traceable to national standards.
Standard Kit Contents the NDT 815 instrument, 1/4" diameter x 5.0 MHz dual element
transducer on 4 foot long cable, (3) size AA Alkaline Batteries, 4 oz. plastic bottle of
Ultrasonic Couplant, Operation Manual, Serial Interface Cable, Data Management Software,
Hard-shell Carrying Case and Certificate of Calibration.
Back
to Top |
MX-5DL
A dual element, pulse-echo gauge for spot check
measurements with data logging.
Small, light weight and rugged, this gauge is very easy to operate and will perform
in a wide variety of situations and conditions.
SPECIFICATIONS
Principle: Ultrasonic Pulse-echo with dual element transducer
Range: 0.030 (0.75mm) to 19.99 (199.99mm) depending upon transducer used and material
Resolution: 0.001 (0.01mm)
Sound Velocity Range: 0.0492 to 0.3930 in/ms (1250 to 10,000 m/s)
Automatic linearity adjustment
Automatic Zero adjustment with transducer on calibration block
Automatic Shut-Off 2 minutes after last reading or last key press
RS232 output to a serial printer, a computer or other external storage device
Automatic numeric Data Logging system for up to 1000 readings
Built in Probe Zero calibration block
Easy to read back lighted Liquid Crystal Display with Coupling Indicator
Instant switching between English or Metric measurement
Up to 200 hours continuous operation using alkaline batteries
Low Battery Indicator
Size: 1.7" H. x 2.4" W. x 4.75" L (32mm x 62mm x 120mm)
Temperature Range: - 20 to 120 F (-20 to 50 C)
Weight: 10 oz. with batteries (284 grams)
Powered by two AA 1.5V alkaline batteries
2 year warranty
Standard kit Includes the MX-5DL with alkaline batteries, 1/4"
diameter x 5.0 MHz dua;l transducer on 4 foot long side-exiting cable, 4 oz. bottle of
ultrasonic gel couplant, Windows based software, PC interface cable, Operation Manual and
foam lined plastic case.
Click for MX-5DL .pdf document
Back to
Top
MMX-6 A dual element gauge with both pulse-echo and multi-echo
(thru paint) measurement capability.
MMX-6DL includes a 1000 reading data logger
Click for MMX-6/6DL .pdf document
SPECIFICATIONS
Physical
Weight: 10 ounces (with batteries).
Size: 2.5W x 4.5H x 1.24D inches (63.5W x 114.3H x 31.5D mm).
Operating Temp: 20° to 120°F (-30° to 50°C).
Case: Extruded aluminum body with nickel-plated aluminum end caps (gasket sealed).
Keypad: Sealed membrane that is resistant to both water and petroleum products.
Nine tactile-feedback keys.
Transducer: Dual-element (transmit and receive). 1 to 10 MHz frequency
Special high-damped transducers available for Through Paint and Coating measurements
Locking quick disconnect LEMO connectors on 4 foot cable.
Custom transducers available for special applications.
Measuring
Resolution: 0.001 inch (0.01 millimeter)
Velocity Range: 0.0492 to .3937 in/µs 1250 to 10,000 meters/second)
Range: Pulse-Echo Mode (Pit & Flaw Detection) measures from 0.025 to 19.999 inches
(0.63 to 500 millimeters).
Echo-Echo Mode (Through Paint & Coatings) measures from 0.1 to 1 Inch (2.54 to 25.4
millimeters).
Switchable to either English or Metric units.
Built in Stainless steel reference disk for probe zeroing.
Measurement Rate: Four readings per second for single point measurements or 16 per second
in Scan Mode-captures the minimum thickness. Calibration: Single and two point calibration
option included.
Features
Multi-Mode:Toggle between Pit & Flaw Detection (Pulse Echo Mode), or Through Paint
& Coatings (Echo-Echo Mode) with a single button.
Alarm Mode: Enter a minimum acceptable thickness value. If measurement falls below minimum
a red LED will illuminate and beeper will sound.
If measurement is above minimum value, green LED will illuminate.
Data Output: RS232 output sends data to a serial printer, a computer or other external
storage device.
Power Source
Two 1.5V alkaline or 1.2V NiCad AA cells.
Typical operation 200 hours on alkaline; 130 hours on NiCad.
Display flashes when battery is low. Unit turns off automatically when battery is too low
to operate reliably.
Display
Multi-function 4.5 digit LCD display with 0.500 inch high numerals, backlit for use in
poor light conditions.
Backlight is selectable on/off/auto (illuminates only when taking a measurement).
Measurements displayed in inches, inches/microsecond, millimeters and meters/second.
Bar graph indicates stability of reading.
Internal Data Logger (MMX-6DL only)
Automatic numeric Data Logging system 1 to 1000 readings.
OBST indicates no reading.
Certification
Factory calibration traceable to national standards.
2 year limited warranty
Standard Kit includes MMX-6 Gauge, 5 MHz dual transducer & cable, 4 oz. bottle of
couplant, operation manual and carry case. PC interface cable and software included with
the DL model.
Back to Top
MMX-7 A dual element multi-echo (thru paint)
gauge with large digit numeric and B-Scan displays and Data Logger
Click for MMX-7 .pdf
document
Pulse-echo Mode for flaw and pit measurement and Multi-Mode
(Echo-Echo mode) to measure through and eliminate paint and coatings. Built-in
automatic gain control.
Time based B-Scan feature displays a cross section of the opposite surface
allowing inspectors to the contour of the back surface.
Selectable low, medium, and high multiple gain settings offer the inspector
the additional punch power for materials that are hard to penetrate.
Store up to 64 custom setups. Factory setups are included for common
applications.
Alpha-numeric data logger to provides increased versatility for custom
reporting needs and stores up to 12,000 measurements.
Built-in transducer types offer increased linearity between transducer
frequencies and diameters.
High speed scan feature provides a better representation during the scanning
process, by making 32 measurements per second. Remove the transducer from the test
material to display the thinnest measurement found during strip scan.
Use the visual alarm to set Hi and Lo limits to alert inspector when the
measurements are out of tolerance.
Windows PC software to transfer data to and from a PC, view B-Scans and general
measurements, set break-points, view statistics, and export files into excel.
2 year limited warranty
Specifications:
Physical
Weight: 13.5 ounces (with batteries).
Size: 2.5 W x 6.5 H x 1.24 D inches (63.5 W x 165 H x 31.5 D mm).
Operating Temperature: -14° to 140°F (-10° to 60°C).
Keyboard: Membrane switch with twelve tactile keys.
Case: Extruded aluminum body with nickel-plated aluminum end caps (gasket sealed).
Data Output: Bi-directional RS232 serial port. Windows® PC interface software.
Display: 1/8 in. VGA grayscale display (240 x 160 pixels). Viewable area 2.4 x 1.8 in. (62
x 45.7mm).
EL backlit (on/off/auto).
Ultrasonic
Measurement Modes: Pulse-Echo (flaws, pits) and Echo-Echo (thru-paint)
Square wave pulser.
Receiver: Selectable low, medium or high gain in pulse-echo mode. AGC gain
control in ech0-echo mode.
Timing: 20 MHz with ultra low power 8 bit digitizer.
Power
Source
Three 1.5V alkaline or 1.2V NiCad AA cells. Typically operates for 150 hours on alkaline
and 100 hours on NiCad (charger not included.)
Auto power off if idle 5 min. Battery status icon.
Measuring
Pulse-Echo Mode Range: (Pit&Flaw Detection) measures from 0.025 to 9.999 inches
(0.63 to 254 millimeters).
Echo-Echo Mode Range: (Thru Paint & Coatings) measures from 0.050 to 4.0 inches
(1.27 to 102 millimeters).
Range will vary +/- depending on the thickness of coating.
Resolution: +/- .001 inches (0.01 mm)
Velocity Range: .0492 to .3936 in/ms 1250 to 9999 meters/sec
Calibration: Single and Two point calibration option, or selection of basic material
types.
Units: English & Metric
Display
Display Views:
Large Digits is Standard thickness view. Digit Height: 0.400 in (10mm).
B-Scan Cross sectional view. Display speed of 15 secs per screen.
Scan Bar Thickness 6 readings per second. Viewable in B-Scan and Large Digit views.
Repeatability Bar Graph Bar graph indicates stability of reading
Data Logger (Internal): 12000 and B-Scans (alpha numeric storage).
Memory: 16 megabit non-volatile ram.
Transducer
Transducer Types: Dual Element (1 to 10 MHz).
Locking quick disconnect “ 00” LEMO connectors. Standard 4 foot cable.
Custom transducers and cable lengths available for special applications.
Features
64 custom user-definable setups. Factory setups can also be edited by the user.
Selectable transducer types with built-in dual path error correction for improved
linearity.
Alarm Mode: Set HI and LO tolerances with audible beeper and visual LEDs.
Fast-Scan Mode takes 32 readings per second and displays the minimum reading found when
the transducer is
removed. Display continuously updates while scanning.
Certification
Factory calibration: traceable to national standards.
Standard Kit Contents: MMX-7 instrument, 1/4" diameter x 5.0 MHz dual element
transducer on 4 foot long cable, (3) size AA Alkaline Batteries, 4 oz. plastic bottle of
Ultrasonic Couplant, Operation Manual, Serial Interface Cable, Data Management Software,
Hard-shell Carrying Case and Certificate of Calibration.
Back to Top
CMX Versatile ultrasonic
thickness gauges that combine the ability to measure wall thickness with or without
coatings and simultaneously measure the coating itself. Time-based B-scan and large
data storage capacity compliment the advanced features of this brand-new gauge.
Available in three models. Standard (CMX), with data logging (CMXDL)
and with data logging, A-Scan and B-Scan display with three independent gates (CMXDL+)

Click for CMX Series
.pdf document
Measurement Modes:
* Coating on: Pulse-Echo Coating (PECT)
* Coating off: Pulse-Echo (P-E)
* Temp Comp: Pulse-Echo Temperature compensated (PETP)
* Thru-Paint: Echo-Echo (E-E)
* Coating Only: Coating (CT)
Features:
* Auto Probe Recognition
* Auto Probe Zero
* Auto Temperature Compensation
* Adjustable Gain - (v-low, low, med, high, v-high)
* Time based B-Scan (cross section view)
* High Speed Scan
* High/Low Alarm - dynamic update of both
* Differential Mode
* Transducer Linearity Selector
* Material Chart
* 64 Custom Setup Capability
* Massive Data Storage (CMXDL & CMXDL+)
(Graphics ON) 16,000+ readings, B-Scans, & gauge settings for every stored reading
(Graphics OFF) 210,000+ readings - including material, coating, min & max for all
readings
CMXDL+ with 2 display options
ELECTRONIC PLATFORM:
Powered by a 100MHz DSP platform using FPGA technology.
Two Channels - Dual pulsers and receivers.
Up to 140Hz pulse repetition rate.
Adjustable gain settings - vlow, low, med, hi, vhi.
Automatic gain control (AGC).
Time corrected gain (TCG).
Massive data storage (32 Megabit of non-volatile RAM).
1/8 VGA Grayscle Display (240 x 160 pixels), EL backlit (On/Off/Auto), update rate of 25
times per second.
1/4 VGA AMOLED Color Display (320 x 240 pixels), 125 Hz refresh rate.
FEATURES:
Viewing Options:
A-Scan (RF waveform, +/- Rectified waveform), Time-Based B-Scan,
and Large Digits.
Measurement modes:
Pulse-Echo, Pulse-Echo w/Coating, Pulse-Echo w/Temperature
Compensation,
Echo-Echo, Echo-Echo Verify, & Coating Only.
Automatic: probe zero, probe recognition, and temperature compensation.
Stores up to 64 custom setups for specific applications.
High Speed Scan of up to 50 readings per second.
Audible/visual alarm with hi and lo limit settings.
Built-in differential mode for QC inspections.
Time based B-Scan feature for cross section material scans.
Attenuator: 20db cut/boost.
pulser: 50 volt cut/boost (100 - 200 volts).
Data storage formats: Alpha numeric grid and sequential w/auto identifier.
Windows PC software included.
2 year limited warranty.
Certification: Factory calibration: traceable to national standards.
Standard Kit Contents the CVX instrument, transducer on 4
foot long cable, (3) size AA Alkaline Batteries, 4 oz. plastic bottle of Ultrasonic
Couplant, Operation Manual, Serial Interface Cable, Data Management Software, Hard-shell
Carrying Case and Certificate of Calibration.
Back to Top
UMX-2 Underwater Metal & Coating Thickness Gauge

The UMX-2 Underwater gauge is the latest addition to the Dakota product range.
Fully equipped with our DSP platform using FPGA technology, it's extremely versatile for
all those offshore inspections. The enclosure is extremely durable, with a maximum depth
rating of 1000ft (300m), and offers the use of both Dual & Single Membrane transducers
to accomodate your application requirements. The features include: auto probe zero, auto
probe recognition, and multiple measurement mode options.
Dual Element modes (corrosion inspections) include: Pulse-Echo (pits & flaws),
Echo-Echo (thru-paint) & Pulse-Echo w/coating (combo). The Single Membrane probe
option uses a Triple-Echo (TCG) mode for coated and uncoated general applications. While
not suitable for blind surface corossion inspections, this option offers good linearity
for general purpose applications.
The calibration section contains 8 common pre-programmed material types and 1 custom user
programable option. Connect the UMX-2 to your PC, and program it with our DakView3 utility
software.
The UMX-2 stores up to 5,000 readings in a variety of configurations. The file structure
options include: sequential with auto identifier and alpha numeric grid formats. Every
measurement stored also saves a copy of the UMX-2 current settings, along with an A-Scan
waveform graphic.
ELECTRONIC PLATFORM:
Powered by a 100MHz DSP platform using FPGA technology.
Two Channels - Dual pulsers and receivers.
Up to 140Hz pulse repetition rate.
Display update rate of 25 times per second.
Adjustable gain settings - vlow, low, med, hi, vhi.
Automatic gain control (AGC).
Time corrected gain (TCG).
Massive data storage (32 Megabit of non-volatile RAM).
FEATURES:
Measurement modes: Dual Element - Pulse-Echo, Pulse-Echo w/Coating, Echo-Echo.
Single Element - Triple-Echo (TCG).
Automatic probe zero and probe recognition.
Data storage formats: Alpha numeric grid and sequential w/auto identifier.
Windows PC software and interface cable included.
3 year limited warranty.
Click for UMX-2 .pdf document
Back to Top
|